Agilent Headspace 5975C Series MSD

We have the Agilent 5975C Series GC/MSD with the New Triple-Axis Detector. It is combined with 7890A GS System CTC Analytics Combi PAL Sample Injectors for liquid injection, headspace, and solid-phase microextraction.

This device has a better MS resolution and the lowest mass deviation, superior sensitivity in all ionization modes and unquestioned spectral integrity establish a solid foundation of consistent quality.

  • Inert ion source (programmable up to 350 °C) delivers enhanced response for active compounds and late eluters
  • True hyperbolic quadrupole provides maximum transmission and resolution
  • Thermally stable quartz quadrupole (can be heated up to 200 °C for long-term mass axis stability)
  • The Triple-Axis Detector with the triple-channel electron multiplier for increased signal and reduced neutral noise
  • Longer lasting tune and calibration stability 
  • Automatic EPC reagent gas adjustment and CI tuning
  • Electron impact (EI) ionization standard with CI ion source from the same source
  • Fast SIM – dwell time as short as 1 ms
  • 1050 u mass range – extends application space for high mass compounds
  • SemiQuant capability for easy and fast estimation of amount of unknown compounds

Agilent’s 5975C Series GC/MSD includes performance electronics to maximize signal transmission for faster GC/MS in full scan, Selected Ion Monitoring (SIM) and Synchronous SIM/Scan modes

  • Scan rates up to 12,500 u/s – compatible with fast full scan GC/MS
  • Run Synchronous SIM/Scan without compromising performance
  • More data points per peak for better peak integration (software includes MSO ChemStation E02.00.493)